XDR Sensor holders
XDR Complete Positioner & Biteblock Kit: Contents and Sterilization Instructions
An XDR Complete Positioner and Biteblock Kit includes:
- 1 - XDR Paralleling Guide System (ShaRing & Litening Rod set ANT/POST/BW )
- 1 - XDR H Stick Guide System (Ring, Rod, H1 and H2 Stick)
-
A pack of the following biteblocks:
-
Yellow Posterior PA Biteblock-Set of 2
-
Blue Anterior PA Biteblock-Set of 2,
-
Red Horizontal BW Biteblock-Set of 2
-
Red Vertical BW Biteblock-Set of 2.
-

XDR Paralleling Guide System (ShaRing & Litening Rod set ANT/POST/BW )
XDR H Stick Guide System (Ring, H1, Rod, and H2 Stick)
Biteblocks- Yellow Posterior PA, Blue Anterior PA , Red Vertical BW, and Red Horizontal BW
Steam Autoclave Guidelines
Temperature |
Pressure |
Time |
Unwrapped Items |
||
|
15 psi (103.4 kPa) |
15 min. |
122° C (250° F) |
30 psi (206.8 kPa) |
3 min. |
Lightly Wrapped Items |
||
121° C (250° F) |
15 psi (103.4 kPa) |
20 min. |
132° C (270° F) |
30 psi (206.8 kPa) |
8 min. |
Heavily Wrapped Items |
||
121° C (250° F) |
15 psi (103.4 kPa) |
20 min. |
132° C (270° F) |
30 psi (206.8 kPa) |
10 min. |
Autoclaving Instructions:
- Autoclave sensor holders before and after each use.
- Carefully follow all Steam Autoclave Guideline instructions above, to prevent damage to instruments, the autoclave, or other items.
- Place components in sterilization pouches: metal and plastic parts separately.
- Position pouches paper-side down on the middle tray, away from walls and the heating element.
- Follow the autoclave manufacturer's instructions for cycle parameters, and do not exceed 270°F (132°C).
Note: Plastic parts have a limited life and should be replaced periodically. Any method of sterilization will shorten the life of plastic parts.
To order XDR sensor holders click here!